Laboratory Instrument
Products Definition
1. Laboratory Magnetic Stirrer:
Laboratory magnetic stirrer is a laboratory instrument used for liquid mixing. It primarily serves the purpose of stirring or simultaneously stirring and heating low-viscosity liquids or solid-liquid mixtures.
2. Laboratory Heating Mantle
Laboratory heating mantle, also known as a heating jacket or heating sleeve, is a device used in chemistry and other scientific fields to heat containers, such as flasks or beakers, uniformly and safely. It typically consists of a fabric or silicone-based outer shell that wraps around the container, with heating elements embedded within the shell.
ACHIEVE CHEM supply the magnetic stirrer and heating mantle below:
Click the laboratory instrument below to get the PRICE LIST
Purchasing Guide
1. Preparing the necessary parameter of your request.
1) How much capacity?
2) Speed Range?
3) Display Mode(LCD)? Yes/No
2. Referring the parameters of the available and on-sale standard specification.
We provide all of the specifications online, you can check all of the details from our official website, download the price list and inquire the sales for one-for-one service.
3. Judgement
1) If the on-sale specification can meet your request, it is economic and stable.
2) If the on-sale specification can't meet your request, we will establish Business-Technology group serve for you for customization project.
Category
Laboratory Magnetic Stirrer:
1) Magnetic Stirrer
b. LCD/Double Knob/Timmer/Heating Plate
c. Stirring Volume: 5L
2) Large Capacity Magnetic Stirrer
a. Double Knob/DC Brushless Motor/100~240V/5~40°C 80%RH
b. Stirring Volume: 10L/20L/50L
3) Multi Link Magnetic Stirrer
a. Double Knob/LCD/100~240V/100~1500RPM
b. Stirring Volume: 3*1/6*1/9*1
4) Mini Magnetic Stirrer
a. Miniature Brushless Motor/Stepless Speed Regulation/0~2000RPM/AC 220V 50Hz
b. Stirring Volume: 2L
5) Single Control Multi Connection Magnetic Stirrer
a. LED Digital/0~1600RPM/RT±5~99.9°C/220V 50/60Hz
b. Stirring Volume:4*1/6*1
https://www.achievechem.com/chemical-equipment/laboratory-magnetic-stirrer.html
Laboratory Heating Mantle:
1) Normal Temperature Regulation
Feature: Heating Function
Capacity: 100ML / 250ML / 500ML / 1,000ML / 2,000ML / 3,000ML / 5,000ML / 10,000ML / 20,000ML
2) Digital Temperature Control Type
Feature: Heating Function, Temperature Display, Temperature Sensor
Capacity: 100ML / 250ML / 500ML / 1,000ML / 2,000ML / 3,000ML / 5,000ML / 10,000ML / 20,000ML / 50,000ML
3) Digital Magnetic Type
Feature: Heating Function, Temperature Display, Temperature Sensor, Magnetic Stirring
Capacity: 100ML / 250ML / 500ML / 1,000ML / 2,000ML / 3,000ML / 5,000ML / 10,000ML / 20,000ML / 50,000ML
4) Digital Display Constant Temperature Magnetic Force
Feature: Heating Function, Temperature Display, Time Display, Temperature Sensor, Magnetic Stirring, Timer
Capacity: 100ML / 250ML / 500ML / 1,000ML
https://www.achievechem.com/chemical-equipment/laboratory-heating-mantle.html
FAQ

01. Laboratory Magnetic Stirrer
1. Mixer does not start or the speed is unstable.
Cause: It may be a power failure, a damaged motor, or a misalignment between the stirrer and the magnetometer.
The solution: Ensure a stable power supply, check the motor status regularly, and adjust the magnetometer position to ensure alignment.
2. The mixer falls off or does not work.
Cause: The connection between the agitator and the magneton may be loose, or the magnetic properties of the magneton may be weakened.
Solution: Check the connection between agitator and magneton regularly to ensure tight; Replace the magnetometer if necessary.
3. Mixer noise is too high.
Cause: May be bearing wear, motor abnormalities, or stirrer and container friction.
Solution: Check bearing and motor status to ensure proper clearance between mixer and container.
4. Stirrer overheating.
Cause: the motor may be overloaded, or poor heat dissipation.
Solution: Make sure the magnetic stirrer works within its rated load, keep the surrounding air well ventilated, and clean the radiator regularly.
5. Agitator gear is broken or stuck.
Cause: the gear may be worn, or foreign objects may enter the gear box.
Solution: Check the gear status regularly and replace the gear with serious wear; Keep the work area clean.
6. Agitator fluid leakage.
Cause: Seals may age, or containers may be damaged.
Solution: Replace aging seals and use intact containers.
02. Laboratory Heating Mantle
1. The first time the electric heater is used, white smoke is emitted.
Cause: This is a normal phenomenon, because the glass fiber may contain oil and other compounds during production.
Solution: Store in a ventilated place and use normally after a few minutes.
2. The heating jacket is not evenly heated.
Cause: It may be that the heating element inside the electric heating sleeve is unevenly distributed, or the container does not match the electric heating sleeve.
Solution: Select the appropriate container and ensure that the container matches the heating sleeve; Check the distribution of heating elements inside the heating sleeve.
3. The temperature control of the heating sleeve is abnormal.
Cause: It may be that the heating system parameters are set improperly, or the heating element is aging and damaged.
Solution: Check and adjust heating system parameters and replace aging or damaged heating elements.
4. Heating mantle overheating.
Cause: the electric heating sleeve may be overloaded for a long time, or poor heat dissipation.
Solution: Ensure that the electric heating sleeve works within the rated load to avoid long-term continuous use; Keep the area well ventilated.
Application

1. Chemical laboratory: It is used in solution preparation, chemical reaction mixing and other processes.
2. Biological laboratory: It is often used in cell culture, microbial culture and other processes.
3. Food industry: It is used in the production of condiments, beverages, dairy products, etc. By stirring, you can ensure that the raw materials are evenly mixed, thereby improving the taste and quality of theproduct.
4. Pharmaceutical industry: It is used for the preparation of drugs, antibiotics, vaccines, etc. Its precise mixing speed and temperature control help ensure the quality and stability of the drug.
5. Environmental monitoring: It is often used in the treatment and analysis of water samples, soil samples and other samples to ensure the uniform mixing of samples and provide accurate data for subsequent testing and analysis.
1. Chemical experiments: In organic chemistry experiments, laboratory heating sleeves are often used to heat chemical reagents to promote chemical reactions.
2. Material synthesis: In material synthesis experiments, it can help obtain high-quality materials by adjusting the temperature to meet the temperature requirements of different materials.
3. Biological experiments: In the biological field, laboratory heating sets are often used to heat cells, proteins, etc., providing more accurate results for biological experiments.
4. Physical and environmental experiments: In physical experiments, laboratory heating sleeves canbe used for thermal analysis of materials; In environmental experiments, it can be used to heat sewage.

Certificate
ACHIEVE CHEM has passed ISO quality system certification, CE certification, UL certification, and is a qualified supplier to multiple Fortune Global 500 companies.
Our reactors sell well worldwide and are widely used in research institutions, universities, and manufacturing industries. For detailed information, please refer to the engineering case later.
In addition, our company also provides technology support services, including new organic materials, senior intermediates and pharmaceutical intermediates.
Welcome to cooperate with ACHIEVE CHEM! We will introduce it later in the engineering case.

![]() |
![]() |
![]() |
![]() |